Appendix 7
Calibration Procedures For pH Electrodesa
app7phel.doc 09/09/98
This calibration procedure should be performed:
o at the beginning of any experiment requiring pH measurements
o whenever switching to another pH electrode after the experiment
has begun, or
o before continuing whenever the meter is turned off
A. Chemicals
"pH 4" buffer solution (the exact pH = pH1)
"pH 7" buffer solution (the exact pH = pH2)
NOTE: buffers of these specific pH values are required to calibrate the Model
F31 or F32 meters in this range. To calibrate other pH meters, or for other ranges, use other appropriate buffers.B. Equipment
0.5 inch magnetic stir bar
magnetic stirrer
pH electrode (determine I.D. No. for electrode used)
pH meter (Model
F31 or F32)C. Procedures
1. Use a thermometer to determine that both buffer solutions are at the same temperature, ambient. Do not proceed until the solutions are at the same temperature (within 0.2oC.). Record the temperatures.
2. Rinse the pH electrode with distilled water and shake gently to remove excess water droplets. Do not wipe the electrode sensing area with Kimwipe or tissue.
3. Totally immerse the electrode sensing area into the "pH 4" buffer solution. FOR THE
F31, turn on the pH meter by pressing "clr" button. Calibrate the pH meter by pressing the graph-like calibration button ("I/ "). Stir (or swirl) the solution. Record the pH reading after the auto-read stops.FOR THE
F32, press "" , then "C" to turn on and clear the meter. Press "pH" then "ISTD/" for calibration. Stir (or swirl) the solution. Record the pH reading after the auto-read stops.4. Replace the "pH 4" buffer solution with the "pH 7" buffer solution.
____________________________________________________________________
(a) G. Mattock, Glass electrodes, John Wiley & Sons, 1965.
A7.1
5. Rinse the pH electrode with distilled water and shake gently to remove excess water droplets. Do not wipe the electrode sensing area with Kimwipe or tissue.
6. Totally immerse the electrode sensing area into the "pH 7" buffer solution. Stir (or swirl) the solution.
FOR THE
F31, press the graph-like calibration button, ("I/ "); observe the pH readout; and record the pH reading (pH3) just before the auto-read stops.FOR THE
F32, press the "ISTD/" button, observe the pH readout; and record the pH reading (pH3) just before the auto-read stops.7. Calculate the error factor, X.
X = (pH2 - pH3)/(pH2 - pH1)
8. If X is greater than 0.2, the electrode is malfunctioning; consult your instructor. Otherwise, the pH electrode is acceptable for use with corrections applied (see below).
9. Record the calibration data and the error factor, X, into the Calibration Record Book for this particular pH electrode.
D. Corrected pH Measurements
The pH meter (Model
F31 or F32) automatically applies a correction factor to pH measurements made after the preceding calibration procedure has been conducted. However, for pH meters which do not automatically correct, the procedure below should be used:A correction factor should be added to all pH readings taken by a previously- calibrated pH electrode. If pH4 is a pH reading taken during an experiment, then the correction factor, Y, will be added to pH4 to give the real pH value of the solution.
Y = X
·(pH4 - pH1)pH4(Corrected) = Y + pH4(observed)
NOTE: The above correction procedure does not have to be applied to pH measurements made with the Model
F31 or F32 pH meter, once it is calibrated, as it performs this correction automatically. The calibration data are obtained, however, to maintain a documented record of acceptable performance of the pH electrode.A7.2
Click here to view and/or print the Table for this section
Click here to view and/or print the Sample Table for this section